2 Installation and Patient Safety Making Connections to the IntelliVue G1/G5
18
M1019A#K11 1.5 m (M1013-61001)
M1019A#K12 3 m (M1013-61002)
3 Equipotential Grounding Terminal; this is used to connect the gas analyzer to the hospital’s
equipotential grounding system.
4 Gas exhaust. If N
2
O and/or other inhalation anesthetics are used during anesthesia, pollution of
the operating room should be prevented. Once the gas sample has passed through the gas analyzer,
it should either be returned to or removed from the anesthesia circuit.
NOTE In some countries where closed loop functionality is not available, gas must not be returned to the
anesthesia circuit.
5 Fan Filter
CAUTION Combinations of medical equipment with non-medical equipment must comply with IEC 60601-1-1.
Never use a multiple portable socket-outlet or extension cord when combining equipment unless the
socket outlet is supplied specifically for use with that equipment.
Connecting the IntelliVue G1/G5 to AC Mains
The IntelliVue G1/G5 has a wide-range power supply that allows you to operate the monitor from an
AC (alternating current) power source of 100 V to 240 V (± 10%) and 50/60 Hz (± 5%).
WARNING Always use the supplied power cord with the earthed mains plug to connect the monitor to an
earthed AC mains socket. Never adapt the mains plug from the power supply to fit an unearthed AC
mains socket.
Do not use AC mains extension cords or multiple portable socket-outlets. If a multiple portable
socket-outlet without an approved isolation transformer is used, the interruption of its protective
earthing may result in enclosure leakage currents equal to the sum of the individual earth leakage
currents, so exceeding allowable limits.
Do not connect any devices that are not supported as part of a system.
Any non-medical device placed and operated in the patients vicinity must be powered via an
approved isolation transformer that ensures mechanical fixing of the power cords and covering of
any unused power outlets.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips M1013A IntelliVue G1 and is the answer not in the manual?

Philips M1013A IntelliVue G1 Specifications

General IconGeneral
BrandPhilips
ModelM1013A IntelliVue G1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals