1 Introduction Performance Specifications
10
CO
2
Measurement
AWRR derived from CO
2
Waveform
N
2
O Measurement
O
2
Measurement
Anesthetic Agent Measurement
Range: 0 to 76 mmHg
Accuracy: 0.5 vol% or 12% relative, whichever is greater
Resolution: 1 mmHg
Rise-time: 350 msec typical
Range: 0 to 60 rpm
Accuracy: ± 1 rpm
Resolution: 1 rpm
Detection Criteria: adaptive threshold
Range: 0 to 100vol%
Accuracy: 2.0 vol% + 8% relative
Resolution: 1 vol%
Rise-time: 500 msec typical
Range: 5 to 100vol%
Accuracy: ± 3 vol%
Resolution: 1 vol%
Rise-time: 500 msec typical
Agent Range (vol%) Accuracy Resolution Rise Time
Halothane 0 - 8.5 0.15 vol% + 15.0% relative 0.05 < 500
Enflurane 0 - 10.0 0.15 vol% + 15.0% relative 0.05 < 500
Isoflurane 0 - 8.0 0.15 vol% + 15.0% relative 0.05 < 500
Sevoflurane 0 - 10.0 0.15 vol% + 15.0% relative 0.05 < 500
Desflurane 0 - 20.0 0.15 vol% + 15.0% relative 0.05 < 500