5Troubleshooting the Gas
Analyzer
This chapter provides a recommended procedure for locating and identifying faults on the gas analyzer.
It details how to identify hardware problems and how to proceed when measurement related INOPs
occur.
It details how to proceed when errors are flagged for:
Failed calibration checks and procedures
Failed diagnostic checks.
Equipment needed for troubleshooting:
Equipment Philips part # or other recommendation Accuracy
IntelliVue G1/G5 PC cable M1013-61005 / 451261005001 n/a
TSI Flowmeter 453564178121 3 ml/min or better
Digital Barometer Recommended: DRUCK DPI 705, 2bar, absolute 2 mbar or better
Watertrap M1657B / 989803110871 n/a
Sample Tubing M1658A / 989803104671 n/a
Calibration Gas Reservoir Bag M1659A / 989803104681 n/a
Calibration Gas M1662A / 451261001391 n/a
Gas Exhaust Return Line M1655B / 989803145671 n/a