Instructions for Use | Therapy Modes and Controls
Control Modes 19
Control Modes
A/C-PC: Assisted/Control-Pressure Control
DESCRIPTION:
The A/C-PC mode provides pressure-controlled mandatory or assist-control breaths. When the Trigger Type is set to Off, the ventilator triggers and
cycles all breaths. When the Trigger Type is not set to Off, then the ventilator or the patient can trigger a breath, and the ventilator cycles all
breaths.
SETTINGS:
Setting Name
Description
Pressure Control
Inspiratory pressure above PEEP
PEEP
Positive end expiratory pressure
Rise Time
Time required for the ventilator to change from the expiratory pressure setting to the inspiratory
pressure setting when the breath is triggered.
Breath Rate
Minimum rate of breaths per minute
Inspiratory Time
Length of the inspiratory phase
Trigger Type
Auto-Trak (passive circuits only)
Sensitive Auto-Trak (passive circuits only)
Flow Trigger (Passive, Active PAP, Active Flow, or Dual Limb circuits)
Off
Flow Trigger
Sensitivity
This control is available when the trigger type is Flow Trigger. The flow trigger initiates when the
patient’s inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity setting.
Flow Cycle
Sensitivity
This control is available when the trigger type is Flow Trigger.
As flow begins to decrease during inspiration, if the patient flow is less than the flow cycle set point, the
device cycles to expiration.
FiO
2
(optional)
Requires model with oxygen blender
SETTABLE ALARMS:
Circuit disconnect
High tidal volume
Low tidal volume
High minute ventilation
Low minute ventilation
High respiratory rate
Low respiratory rate
ILLUSTRATION
1. EPAP
2. Pressure control