Instructions for Use | Therapy Modes and Controls
Therapy Control Settings 27
Therapy Actions
Oxygen Flush
DESCRIPTION
This feature requires the oxygen blender. When active, the device delivers 100% oxygen for two minutes. This feature functions independent of
any oxygen blending setting. During an oxygen flush, the High FiO
2
alarm is disabled.
WORKING WITH OXYGEN FLUSH
To start an oxygen flush:
Tap 100% O
2
in the status bar and then tap Start. A timer appears that counts down
the two minutes.
To stop an oxygen flush:
Tap 100% O
2
in the status bar and then tap Stop.
APPLICABLE THERAPY MODES
All
Manual Breath
DESCRIPTION
Delivers a breath based on the current therapy mode settings.
To deliver a manual breath:
Tap the Manual Breath button in the status bar and then tap Start.
APPLICABLE THERAPY MODES
A/C-PC
A/C-VC
PSV
S/T
SIMV-PC
SIMV-VC
Therapy Control Settings
Therapy control settings can be interdependent. For guidance, see the previous therapy mode descriptions.
Setting Name
Setting Range/Increment
Backup Pressure
Support
Adult patient type:
All circuits but passive: 0-60 cmH
2
O
Passive circuit: 0-57 cmH
2
O
Increments: 1 cmH
2
O
Pediatric patient type:
All circuits: 0-30 cmH
2
O
Increments: 1 cmH
2
O
Infant patient type: