EasyManua.ls Logo

Philips Trilogy Evo Universal - Page 55

Philips Trilogy Evo Universal
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instructions for Use | Alarms and System Messages
Alarms and System Messages 55
LOW SPO
2
Priority
Medium
Why it occurs
The measured SpO
2
is less than or equal to the Low SpO
2
alarm setting while the device is delivering
therapy or is in standby.
What to do
Ensure that oxygen tubing is attached to the ventilator and the oxygen source.
Ensure that the oxygen source is providing oxygen.
Increase oxygen flow or increase FiO
2
.
Requirements
Requires pulse oximeter
Device
performance
during active
alarm
The alarm is automatically resolved when the measured SpO
2
rises above the Low SpO
2
alarm setting.
Device will continue to operate.
Alarm Settings
50 to 95% in increments of 1%.
HIGH SPO
2
Priority
Medium
Why it occurs
The measured SpO
2
is greater than or equal to the High SpO
2
alarm setting.
What to do
Ensure that the oxygen source is appropriate.
Decrease percentage of oxygen delivery.
Requirements
Requires pulse oximeter
Device
performance
during active
alarm
The alarm is automatically resolved when the measured SpO
2
falls below the High SpO
2
alarm setting.
Device will continue to operate.
Alarm Settings
90 to 100% in increments of 1%.
LOW ETCO
2
Priority
Medium
Why it occurs
The measured EtCO
2
is less than or equal to the Low EtCO
2
alarm setting for 10 seconds while the
device is delivering therapy.
What to do
Check the patient. If there is a tracheal tube, verify it is properly inserted. Verify that leak is not
excessive. High leak will reduce the measured EtCO
2
. Review the ventilation parameters including tidal
volume, minute ventilation and respiratory rate and verify that they are appropriate.
Requirements
CO
2
sensor
Device
performance
during active
alarm
The alarm is automatically resolved when the measured EtCO
2
falls below the High EtCO
2
alarm
setting.
Device will continue to operate.
Alarm Settings
1 to 100 mmHg in increments of 1 mmHg.
HIGH ETCO
2
Priority
Medium
Why it occurs
The measured EtCO
2
is greater than or equal to the high EtCO
2
alarm setting for 10 seconds while the
device is delivering therapy.
What to do
Passive circuit: check for an insufficient leak
Active and dual limb circuits: ensure the valve is operable
Requirements
CO
2
sensor
Device
performance
during active
alarm
The alarm is automatically resolved when the measured EtCO
2
falls below the high EtCO
2
alarm
setting.
Device will continue to operate.
Alarm Settings
1 to 100 mmHg in increments of 1 mmHg.

Table of Contents

Related product manuals