Instructions for Use | Alarms and System Messages
Alarms and System Messages 58
The alarm is medium priority and the icons are
yellow when 25 minutes of time remains.
The alarm is high priority and the icons are red
when 15 minutes of time remains.
Switch to alternate battery or AC power to recharge the low battery. If the low battery is recharged
and the alarm continues, replace the battery.
performance
during active
Device will continue to operate.
AC DISCONNECTED
AC power is disconnected while the ventilator is providing therapy or in standby using AC power.
Switch to alternate power source.
performance
during active
Device will continue to operate.
INTERNAL BATTERY IN USE
The power source has switched to the internal battery.
You start therapy on internal battery power.
Confirm the remaining battery capacity. This is the last available power source.
• Prepare an alternative power source such as AC power or an external battery.
performance
during active
Device will continue to operate.
REPLACE DETACHABLE BATTERY
The detachable battery has failed or the battery state of health is less than or equal to 60%.
Replace the detachable battery.
• If the alarm continues, plug the device into an alternate AC power source.
performance
during active
Device will continue to operate.
If the condition still exists 60 minutes after you manually reset the alarm, the alarm repeats.
INTERNAL BATTERY DEPLETED
The internal battery is depleted.
Plug the device into an alternate AC power source, or connect to a fully charged detachable or
external battery.
performance
during active
Device will continue to operate.