Instructions for Use | Technical Specifications
Specifications 89
Relative humidity: 5% to 90% RH, non-condensing
Atmospheric pressure: 62 to 106 kPa
Altitude: -1261 to 12,971 feet
Battery charging temperature: 5°C to 40°C
Transient operating temperature,
excluding high pressure oxygen
blending
5% to 93% RH, non-condensing
19.3 cm D x 28.6 cm W x 24.5 cm H
7.6” D x 11.25” W x 9.65” H
IP22: protection against finger-sized objects and protected against
dripping water when tilted up to 15 degrees.
IEC 60601-1 classification
Type of protection against electric shock: Class II equipment
Degree of protection against electric shock – type BF applied part
This device does not contain natural latex rubber or dry natural rubber.
100V – 240V - 50/60 Hz 1.7A
Internal and detachable Li-ion
batteries
7.5 hours nominal total run time per method in IEC 80601-2-72 (each
battery)
Charge time for detachable and internal battery
from 0% to 80%: 2.5 hours
from 0% to 100%: 3.5 hours
Degree of protection against electric
shock
Alarm sound pressure level
Lowest setting: ≥ 53.9 dBA
Highest setting: ≥ 85.5 dBA
Sound pressure level - ventilator,
circuit and exhalation device
43.7 dBA as measured according to ISO 80601-2-12
Sound power level -ventilator, circuit
and exhalation device
51.6 dBA as measured according to ISO 80601-2-12
0 to 30 l/min; maximum 10 psi
280 to 600 kPa (41 to 87 psi)
Ventilator response to a 21 -90%
increase in oxygen concentration
±(2 cmH
2
O + 4% of the setting)