Instructions for Use | Technical Specifications
EMC Information 95
Electromagnetic Environment - Guidance
Electromagnetic
compatibility
(EMC). Part 4-8:
Testing and
measurement
techniques. Power
frequency (50/60
Hz) magnetic field
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical hospital
or home environment.
NOTE: UT is the AC mains voltage prior to application of the test level.
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it
is used in such an environment.
Electromagnetic Environment - Guidance
-6
Electromagnetic
compatibility (EMC). Part
4
-6: Testing and
measurement techniques.
Immunity to conducted
disturbances, induced by
radio-frequency fields.
Portable and mobile RF communications
equipment should be used no closer to any part of
the device, including cables, than the
recommended 30 cm separation distance.
6 Vrms
Amateur Radio &
ISM
Bands between
6 Vrms
Amateur Radio & ISM
Bands between 150 kHz
and 80 MHz
-4-3
- Part
-3: Testing and
Radiated, radio-
frequency,
electromagnetic field
immunity test
Up to 28V/m in
telecommunication
bands as specified in
clause 8.10 of IEC
60601
-1-2
Up to 28V/m in
telecommunication bands
as specified in clause 8.10
of IEC 60601
-1-2