LIFEPAK CR Plus and LIFEPAK EXPRESS Defibrillator Operating Instructions 5-5
©2009-2014 Physio-Control, Inc.
5 Caring for Your Defibrillator
Caring for Your Defibrillator
To discharge and dispose of a used battery charger:
1 Insert the discharger into the used battery
charger.
Note: Do not attempt to remove the discharger
once it is in place.
2 Let the discharger fully deplete the battery
charger. Wait at least 9 days.
3 Recycle the battery according to national and
local regulations. Refer to page 5-8 for more
information.