Maintenance and infection control
After every patient
Surfaces
1 Wipe with Planmeca approved surface
disinfectant:
• metal parts of headrest
• metal parts of armrests
• protective fascia of the operating light
• handles of the operating light
• hygienic membrane
• instrument hoses
• instruments (see instrument instructions)
• instrument console*
• balanced instrument arms
• tray table
• cuspidor top cover
Wipe the patient mirror of the operating light
with a mild soap- and water solution.
2 Wipe the upholstery with a mild soap- and
water solution. Wipe with Planmeca approved
upholstery disinfectant ifneeded.
3 Wipe the monitor and the handle (seemonitor
instructions).
4 Clean the bowl:
• Pour a few drops of Planmeca approved
suction disinfectant into thebowl and use
asoft brush.
• Rinse the bowl by pressing the bowl
rinsebutton.
• Wipe the outside of the bowl.
3
*Before cleaning the instrument console, lock the touch screen.
Touch screen is unlocked. Press to lock.
Touch screen is locked. Press and hold to unlock.
4
1
2
Do not
spray!
Suction system
1 Remove used aspirating tips.
2 Rinse the suction tubes by aspirating one
glass ofwater with each suction tube.
3 Wipe the suction handpieces with Planmeca
approved surface disinfectant.
4 Wipe the Flexy™ holder with Planmeca
approved surface disinfectant.
Do not
spray!
32 4
Waterlines
Short ushing
1 Place all water consuming instruments
intotheushingholder.
2 Bend the instrument arms to an angle of
atleast 90° and start short ushing.