ATTACHMENT A
VERIFICATION TECHNIQUE
Attention! The present verification technique is used for the wrist gamma
dosimeter PM1603, the verification of PM1603A and PM1603B modifications of
PM1603 dosimeter is carried out according to this verification technique. The
verification should be carried out in accordance with the items of verification
technique which are relevant to the present dosimeter modification.
А.1 Introduction
This verification technique extends to wrist gamma dosimeters PM1603, PM1603A
PM1603B, corresponds to the Methodical Instructions 1788 "Radiation-monitoring devices
for measuring the exposure dose and exposure dose rate, absorbed dose and absorbed
dose rate of photon radiation in air. Verification technique" and establishes the verification
technique for dosimeters.
The verification should be carried out by local bodies of the metrological department
of the State Standard Committee and by institutions authorized to carry out these works.
The verification of a dosimeter should be carried out when releasing for sale, when
releasing from repair and during operation and storage with a time interval of 12 months.
А.2 Operations and measuring instruments
The operations that should be performed during the verification procedure and
measuring instruments that should be used are listed in Table А.1.
Table А.1
Operations
Technique
section
Names of reference and auxiliary measuring
instruments and major performances
External
examination
А.7.1
-
Testing:
А.7.2
-
Determination
of metrological
characteristics
А.7.3,
А.7.4
Dosimetric verification assembly with source
137
Cs
according to Methodical Instructions 2050-90. The
maximum certified error of the assembly should be
not higher than 5 % at 0.95 confidence level
-
А.5
Barometer with the least division of 1kPa.
Measurement range from 60 to 120 kPa
-
А.5
Thermometer with the least division of 0.1°C.
Measurement range from 10 to 30 °C
-
А.5
Hygrometer with measurement range from 30 to 90 %
-
А.5
Stopwatch with measurement range from 1 to 600 s
-
А.5
Dosimeter DBG-06T. Intrinsic error is ±15% (Other
dosimeters providing the required accuracy of
measurements may be used).