INITIAL SET-UP
The device is designed for simple installation. To install, follow the subsequent procedure.
UNPACKING THE DEVICE
Unpack the device and its components and verify that the components are intact and present. It is
recommended to save the packaging box and materials for future transport.
Check parts according to the kit content list.
NOTE: Check that the system and all its components are intact. If you detect any sign of damage
or a missing item, do not continue with the installation. Immediately report this to your Pollogen
representative.
GENERAL INSTALLATION REQUIREMENTS
Electrical Requirements
The System has a universal power supply Single phase 100 - 240 VAC; 50-60Hz. It must be used
only with the supplied power supply.
Environmental Requirements
Corrosive materials can damage electronic parts; therefore, the System should only be operated
in a non-corrosive atmosphere. For optimal operation of the System, maintain room temperature
between 5˚C to 30˚C, keep relative humidity at less than 80%.
For optimal storage, maintain temperature between -20˚C to 60˚C, keep relative humidity at less
than 90% and maintain Atmospheric pressure of 50kPa to 106kPa.
PREPARATION FOR OPERATION
Reminder: Before starting the installation, disconnect the power supply cord.