110-511 PSC/EN 21
Technical Information
Performance Specifications
Operating Temperature Range: Ambient +10° to 135°C
Ambient +20° to 275°F
Temperature Stability: ±0.07°C / ±0.13°F
Heater Wattage: 1100 watts
Electrical Requirements: 120V, 60Hz, 10 amps or
240V, 50Hz, 6 amps
Environmental Conditions Indoor use only
Maximum Altitude: 2000 meter
Operating Ambient: to 35°C (41° to 95°F)
Relative Humidity: 80%, non-condensing
Installation Category: II
Pollution Degree: 2
Ingress Protection: IP 31
Climate Class: SN
Software Class: B
Output Waveform: Sinusoidal
Specifications subject to change without notice.
Attainable Temperatures:
Your MX Immersion Circulator can be used with reservoirs of various capacities and shapes as well as
different fluids. These variables may adversely affect temperature accuracy and stability. For example, a
reservoir with a large surface area loses heat more quickly, which may prevent the Circulator from
attaining the desired temperature.