PRK-6000 Operation Manual
- v -
CAUTION
Do not position the equipment to make it difficult to operate the
disconnection device. (Appliance coupler or separable plug)
CAUTION
Do not placed the multiple socket-outlet for PRK-6000 system on the
floor in order to prevent liquid penetration and damage to the product.
CAUTION
PRK-6000 system shall not be connected with additional multiple
socket-outlets or extension cords in addition to a designated single
multiple socket-outlet.
CAUTION
Maximum permissible load of each socket-outlet used for the PRK-
6000 system, shall not be less than 100VA.
CAUTION
If non-medical electrical equipments (e.g. Video monitor, IT equipment,
etc.) that are connected with the PRK-6000 are, directly connected to
the wall socket-outlets, high touch current may flow since the earth
continuity is not ensured.
CAUTION
Multiple socket-outlet should be a grounding-type . and complied with
IEC 60884-1.
CAUTION
Connection of the plug shall be possible only by using the tool.
(Refer to the figure below.)