26 2726
EEx / I.S. approval - 5350B:
KEMA 02ATEX1318 ..................................... II 1 GD or II 2 (1) GD,
T65°C...T105°C
EEx ia IIC or EEx ib [ia] IIC T4...T6
Applicable in zone ....................................... 0, 1, 2, 20, 21 or 22
FM, UL and CSA ......................................... IS, Class I, Div. 1, Group A, B, C, D
IS, Class I, Zone 0/1, Group IIC
IS, Class I, Div. 2, Group A, B, C, D
Ex / I.S. data:
Signal output / supply, terminal 1 to 2:
Max. ambient temperature depends on the
P
o
of the connected barrier.
Sensor input, terminal 3, 4, 5 and 6:
U
o
................................................................ : 5.7 VDC
I
o
................................................................. : 8.4 mA
P
o
................................................................ : 12 mW
L
o
................................................................ : 200 mH
C
o
................................................................ : 40 µF
GOST R approval:
Standard: VNIIM .......................................... Cert. no. Ross DK.ME48.V01899
Ex: VNIIFTRI ................................................ Cert. no. Ross DK.GB06.V00100
Ex permit ..................................................... РРС00-17800
Observed authority requirements: Standard:
EMC 2004/108/EC
Emission and immunity ....................... EN 61326-1 : 2006
ATEX 94/9/EC .............................................. EN 50014 : 1997, EN 50020 : 2002,
EN 60079-15 : 2003, EN 50284 : 1999,
EN 60079-27 : 2005
FM ............................................................... 3600, 3610, 3611
UL ................................................................ UL 508, UL 1604, UL 913,
UL 60079-0, UL 60079-1
CSA, CAN / CSA ......................................... C22.2 No. 142, No. 157, No. 213,
CAN / CSA .................................................. E79-0, -11, -15
ANSI / UL .................................................... UL 60079-0, -11, -15
Unit
5350B
Zone 0, EEx ia IIC, Entity / Fisco
IS, Class I, Division 1, Group A, B, C, D
Entity / Fisco
Barrier where
P
o
< 0.84 W
Barrier where
P
o
< 1.3 W
Suitable for
FISCO systems
Suitable for
FISCO systems
U
i
I
i
P
i
L
i
C
i
T1...T4
T5
T6
30 VDC
120 mADC
0.84 W
1 µH
2.0 nF
T
amb.
< 85°C
T
amb.
< 70°C
T
amb.
< 60°C
30 VDC
300 mADC
1.3 W
1 µH
2.0 nF
T
amb.
< 75°C
T
amb.
< 65°C
T
amb.
< 45°C
17.5 VDC
250 mADC
2,0 W
1 µH
2.0 nF
T
amb.
< 85°C
T
amb.
< 60°C
T
amb.
< 45°C
15 VDC
900 mA
5.32 W
1 µH
2.0 nF
T
amb.
< 85°C
T
amb.
< 60°C
T
amb.
< 45°C
Unit
5350B
Zone 1, EEx ib IIC, Entity / Fisco
IS, Class I, Division 2, Group A, B, C, D
Entity / Fisco
Barriere where
Po < 5.32 W
FISCO segment
coupler
U
i
I
i
P
i
L
i
C
i
T1...T4
T5
T6
30 VDC
250 mADC
5.32 W
1 µH
2.0 nF
T
amb.
< 85°C
T
amb.
< 75°C
T
amb.
< 60°C
17.5 VDC
All
All
1 µH
2.0 nF
T
amb.
< 85°C
T
amb.
< 75°C
T
amb.
< 60°C

Related product manuals