■ 5 Operation
42
5.5 Special operating keys
5.5.1 The tare key
– Ensure that there is no drying taking place, i.e. that the instrument
is in weighing mode.
• Activating taring
– Press «T» briefly.
– A taring procedure is performed.
• High resloution mode (HR) selection
(This option is only available on HR models)
– Ensure that the moisture analyser is in the Weighing mode
– Hold down
«T» until „HR MODE ON” or „HR MODE OFF” is dis-
played
– Release «T»
– If „HR MODE OFF” is selected the moisture analyser is working
only in the low resolution mode and therefore the readability is 10
times lower than in the high resolution mode.
• Activating calibration
– Keep «T» pressed until "BALANCE CALIBRATION" is displayed.
– Release «T».
– The balance carries out a calibration in accordance with the
settings in the configuration menu and logs these in the form of a
printout (see chapter 5.3.6 "Balance calibration" and see chapter
9.3.1 "Calibrating the balance").
• Activating a temperature calibration
– Keep «T» pressed until "TEMP. CALIBRATION" is displayed.
– Release «T».
– The moisture analyser carries out a temperature calibration and
logs it in the form of a printout (see chapter 5.3.7 "Temperature
adjustment" and see chapter 9.3.2 "Temperature adjustment").
NOTE
The balance calibration functions is only active if this is switched
on in the configuration menu.
A calibration or an adjustment procedure can be interrupted by
pressing «ON/OFF». This applies both to the balance and to the
temperature.