■ 10 Overview
78
10 Overview
10.1 Technical data
Spezification XM 60 (-HR) XM 66
Heat source, heater type Halogen / Infra-red /
Dark Radiator
Halogen / Infra-red /
Dark Radiator
Weighing range [g] / Readability [g] 124 / 0.001 (0.0001) 310 / 0.001
Drying:
Readability [%] 0.01 (0.001) 0.01
Reproducibility in approx. 1g [%] 0.2 (0.1) 0.2
Reproducibility in approx. 10g [%] 0.02 (0.01) 0.02
Sample weight [g] 0.2 - 124 0.2-310
Result calculations: 100-0%, 0-100%
ATRO 100-999%,
ATRO 0-999%,
G/KG, RESIDUAL, LOSS
100-0%, 0-100%
ATRO 100-999%,
ATRO 0-999%,
G/KG, RESIDUAL,
LOSS
Heating:
Temperature range [°C] / Increment
[°C]
30 - 230 / 1 30 - 230 / 1
Heating methods Standard, Boost, Soft Standard, Boost, Soft
Intervals Boost + 1 Boost + 1
Booster +40% during n.n min
(0.1 - 99.9)
+40% during 3 min
Switch-off criteria:
Autostop [d/s] freely definable
1 - 99 / 10 - 90
5 fixed settings or
freely definable
1 - 99 / 10 - 90
Autostop [%/s] freely definable
0.1 - 99.9 / 10 - 90
-
Adapstop x x
Timer stop [min.] 0.1 - 240.0 0.1 - 240.0
Monitoring:
Viewing window x x
Audio x x
Printout:
GLP x x
Printout - Interval [min.] 0.1 - 10.0 0.1 - 10.0
Sample numbering x x