5 Operation ■
45
5.5.3 The change key
• Changing units
– If drying has already been performed or is under way, the unit of
measurement displayed can be changed using the change key
«
F». Once a drying procedure is ended, the drying result can be
displayed in all the available units until the start of a new drying
sequence using the change key «
F».
– Release «
F» once the unit to which you wish to change is
displayed.
5.5.4 The start/stop key
• Starting drying manually
– Ensure that there is no drying taking place, i.e. that the instrument
is in weighing mode.
– Press «START/STOP» briefly.
– The measurement is started.
• Stopping drying manually
– Drying is under way.
– Press «START/STOP» briefly.
– The measurement is stopped.
• Statistics Info (only XM 60, XM 60-HR)
– Keep «START/STOP» pressed until "STATISTICS INFO" is dis-
played.
– Release «START/STOP»
Method : TEST Boost
Dryer setup :
Heat mode : Boost
Temperature : 140 C
Timer stop : off
Time : 10.0 min
Auto stop : AdaptStop
Free : 1/20 D/s
Autostart : off
Standby Temp : off
Temperature : 40 C
Unit : 100-0%
Weight check : off
Nominal : 3.500 g
Upper limit : 4.000 g
Lower limit : 2.500 g
Settings for the second method
(all the methods saved are printed)
etc....