7 Ash residue program ■
63
Reweighing the ashed samples:
+ 0.000 9
«T»
Tare the device without load
Tare
+ 2.535 9
«T»
Place the empty pan1 onto
the balance. Measure tara 1.
tare pan
+ 4.809 9
«E»
Pour the sample into the pan.
Confirm original weight
original 0.000
+ 4.809 9
original 4.609
Repeat these steps for all samples (back to the weighing mode with «esc»)
Ash the samples externally.
Display Key Step
«START»
Keep the key pressed until
"ASH RESIDUE" is displayed.
Release the key.
- - - - - - -
«A»,
«B»
Select the weighing number 1
(resp. 2-4)
-1- -2- -3- -4-
+ 0.000 9
«D»
Change to residual
original
+ 0,000 9
«E»
Confirm choice
residual
+ 0.000 9
«T»
Tare the device without load
Tare
+ 0.235 9
Put the pan with the ash onto
the balance
residual 0.000 9
Display Key Step