9 Service
72
External calibration with a freely definable weight
"SET CALIBRATION MODE EXT.-DEF." must be set in the
configuration menu (see chapter 5.3.6 "Balance calibration") for
external calibration with a freely definable weight.
You then need to enter the effective value of the calibration weight
(DEF. n,nnnn g) with up to ten times the precision of the instrument
balance.
Procedure to follow:
- - 50 9
Place the calibration weight on the
weighing pan.
The display continous to flash.
+50,000 9
Calibration is complete when the
display stops flashing
NOTE
Is calibrated with the free weight, so only this weight may be used.
Display Key Step
+0,000 9
The balance is in Weighing mode.
+0,000 9
«T»
Press until "BALANCE CALIBRA-
TION" apears.
balance CALIBRATION
- - 0000 9
The balance carries out a Zero
measurement "0000 g" is shown
flashing).
- - 22 9
After the zero measurement the dis-
play flashes with the previously en-
tered calibration weight.
Display Key Step