C960 Series Commercial Treadmill
Page 42
23. If the hand held signal is now being accept, something in the near vicinity is radiating RF
energy that is being received by the chest strap portion of the heart rate PCA. Disabling the
chest strap signal proves that it is radiated energy that is causing the problem.
24. The source of the radiated energy must be determined and relocated so that it no longer
affects the heart rate PCA. Televisions, cell phones, Cardio-theatre receivers, etc. are
possible sources of radiated energy.
25. Set the on/off switch in the “off” position, and remove the spare jumper from J14 of the heart
rate PCA. Re-locate all potential sources of radiation. Set the on/off switch in the “on”
position and repeat the procedure in step 6.