16 www.pridemobility.com J6 with NE, NE+, and Q-Logic Electronics Technical Troubleshooting Guide
Communication or Software Error
Symptoms:
The batteries are fully charged.
All electrical components are properly connected.
The power turns on but displays a fault code other than previously discussed. See table 1, 2, or 3.
A communication or software fault is found under “Actual Faults” or “Fault History” in the diagnosis section of the
handheld programmer. See figures 13 and 14.
Diagnosis:
There is a module error in the system.
Solution:
Use the following procedure to find the source of the error:
1. Ensure that the BUS connections are correct. See diagram 5 or 6.
2. Measure resistance across connectors 3a and 3b of the BUS cable extension (3). See figure 19.
If the multimeter indicates an open for any of the measurements, then replace the BUS cable extension (3) and retest
the system.
If the multimeter indicates less than 1 ohm for all of the measurements, then go to the next step.
3. If the power chair is equipped with multiple input devices, then ensure there is a multiplier in the BUS line.
If there is no multiplier, then contact Quantum Technical Service for further troubleshooting help.
If there is a multiplier on the BUS line, then go to the next step.
4. Measure resistance on the multiplier BUS cable. See figure 20.
If the multimeter indicates an open for any of the measurements, then replace the multiplier and retest the system.
If the multimeter indicates less than 1 ohm for all of the measurements, then contact Quantum Technical Service for
further troubleshooting help.
Open Motor Error
Symptoms:
The batteries are fully charged.
All electrical components are properly connected.
The power turns on but displays “Open Motor” (NE or NE+ controllers) or a motor error (Q-Logic controller). See
table 1, 2, or 3.
One or both motors will not operate.
Diagnosis:
There is a problem with the left motor (6), right motor (7), and/or the wiring between the motor(s) and the power module
(2).
Solution:
Use the following procedure to find the source of the error:
1. Remove the seat. Refer to the power base owner’s manual.
2. Remove the rear and center shrouds. See figure 3.
3. Unplug connector 6a from connector 8b. See diagram 5 or 6.