NORMI 13 (T42023)
Operating Manual - Accessories
26 D703.131.00/03 en
4 Accessories
4.1 Use of the Bucky
Mounting Device T20005
with a Test Object
Assembly
Place the test object with the PMMA absorber
(30 mm) and the bucky mounting device
T20005 on a suitable table.
Place the NORMI 13 test object on the PMMA
absorber. Join both components by means of
four knurled screws.
Undo the screws of the bucky mounting
device.
Then use the two star knob bolts (4, Figure 5)
to fix the PMMA absorber (threaded bores
on the sides) on the bucky mounting device
T20005.
Using a ruler or the edge of the table, align the
height of the test object in the bucky mounting
device.
Set the angle bracket for fitting the bucky
mounting device to the thickness of the wall
bucky.
Tighten the star knob bolts on the bucky
mounting device.
Holding the assembled bucky mounting device
with both hands, suspend it in the wall bucky.
It may prove necessary to adjust the position
of the bucky mounting device.
Disassembly
Holding the assembled bucky mounting device
with both hands, lift it off the wall bucky. To fur-
ther disassemble the bucky mounting device,
place it on a suitable table.
Unscrew the star knob bolts (4, Figure 5) and
remove the test object.