Document No.: 1P0446-1 Rev. B, April 7, 2014
Each flowmeter has a solenoid valve that shuts off the flow of gas, thus retaining the setting of the
respective flowmeter for a subsequent process. For added safety, the gas solenoid for each
flowmeter will not allow gas to flow into the chamber until all interlocks are activated.
Note: The inlet pressure from the gas bottle should be regulated to 10 psig before the gas is
introduced to the back of the IoN 3 MHz.
Table 1 Flowmeter Calibration Data
float material: GLASS BALL float density: 2.55 G/CC
metered Temp: 70ºF metered Pressure: 10 PSIG
tube #A-125-3 units: CC/MIN
PRESSURE INDICATOR
The IoN 3 MHz will report the actual pressure reading (accurate within 20 mTorr) on the front
PRESSURE display. This pressure reading will give a reading of the chamber pressure. After the
PROCESS START button has been depressed, the pressure sensor signals the system logic to start the
process (RF and gas flow) when a crossover pressure of .35 Torr (350mT) has been achieved.
The PRESSURE indicator does not display a reading until reaching a pressure below 2 Torr. The
pressure sensor has a range of around .01 Torr (10 mTorr) to 2 Torr. Normal operating pressure will
range between .5 Torr (500mT) and 1.2 Torr.
DIAGNOSTIC LED DISPLAY INDICATORS
Five LED display indicators are located on the lower front panel of the IoN 3 MHz system. When
illuminated, the LEDs provide the user confirmation that all the system modules are functioning
correctly.