Using eCAL to retrieve the original factory
calibration certicate
The original factory calibration results for the RD7000+ locator can be retrieved from
the unit. Each time the RD7000+ is calibrated either at Radiodetection or an approved
Radiodetection service centre, the calibration results are stored within the locator. To
retrieve the results and print a certicate, carry out the following:
Note: You do not need to purchase an eCAL validation key to retrieve the original
factory calibration certicate.
Register your RD7000+ locator by going to:
www.radiodetection.com/extendedwarranty. See page 3 for more details.
Download Centros Manager by going to:
www.radiodetection.com/centrosmanager. See page 13 for details.
Connect the RD7000+ via the USB connector inside the battery compartment
to a suitable USB port on a PC or laptop.
Switch on the RD7000+ (no segments will be lit but the backlight will be on).
Open Centros Manager and click on Locator eCAL Validation.
Click on Get Original Calibration Data and a message box will automatically pop
up displaying the progress.
In less than 3 minutes the original calibration certicate will be available to view
or print, locate the serial number of the RD7000+ within the Unit Manager window
and expand the contents. Expand Calibration History and click on each date.
When you click on each date, the type of certicate, whether an eCAL validation or
original factory calibration, together with date and serial number will be displayed.
By clicking on this window, the certicate will be displayed and you can choose to
print this certicate.
1.
2.
3.
4.
5.
6.
7.