- 86 -
Response options:
C1_I CR LF
- command understood but not accessible at this moment
C1_A CR LF
- command understood and in progress
MASS FRAME
- response: mass value in a basic measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S I space
stability
marker
space character mass space unit CR LF
Switch off continuous transmission in a basic measuring unit
Format: C0 CR LF
Response options:
C0_I CR LF
- command understood but not accessible at this moment
C0_A CR LF
- command understood and carried out
Switch on continuous transmission in a current measuring unit
Format: CU1 CR LF
Response options:
CU1_I CR LF
- command understood but not accessible at this moment
CU1_A CR LF
- command understood and in progress
MASS FRAME
- response: mass value in a current measuring unit
Response format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S U I
stability
marker
space character mass space unit CR LF
Switch off continuous transmission in a current measuring unit
Format: CU0 CR LF
Response options:
CU0_I CR LF
- command understood but not accessible at this moment
CU0_A CR LF
- command understood and carried out
Set min checkweighing threshold
Format: DH_XXXXX CR LF, where: _ - space, XXXXXmass format
Response options:
DH_OK CR LF
- command carried out
ES CR LF
- command not recognised (mass format incorrect)
Set max checkweighing threshold
Format: UH_XXXXX CR LF, where: _ - space, XXXXX - mass format
Response options:
UH_OK CR LF
- command carried out
ES CR LF
- command not recognised (mass format incorrect)
Give value of min checkweighing threshold
Format: ODH CR LF
Response: DH_MASA CR LF - command carried out
Response format:
1 2 3 4-12 13 14 15 16 17 18 19
D H Space mass space unit space CR LF
Mass
- 9 characters with right justification
Unit
- 3 characters with left justification
Give value of max checkweighing threshold
Format: OUH CR LF
Response: UH_MASA CR LF - command carried out