EasyManua.ls Logo

R&S NRPxxS Series - Page 33

R&S NRPxxS Series
184 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating Concepts
R&S
®
NRPxxS(N)
33User Manual 1177.5079.02 ─ 10
R&S NRPZKU cable or an R&S NRPZ5 sensor hub and an R&S NRPZK6 cable
to connect the sensor to the computer
Windows computer with installed:
R&S NRP Toolkit V 4.17 or higher
R&S NRPV version 3.2 or higher (refer to the operating manual of the R&S
NRPV for a description of the installation process)
Setup
Figure 5-2: Setup with an R&S NRPV
1 = Signal source
2 = R&S NRPxxS(N) power sensor
3 = Host interface connector
4 = R&S NRPZKU cable
5 = USB connector
6 = Computer with installed R&S NRPV
Incorrectly connecting/disconnecting the R&S NRPxxS(N) power sensors can damage
the power sensors or lead to erroneous results.
Ensure that you connect/disconnect your power sensor as described in Chapter 3,
"Preparing for Use", on page 11.
Starting a measurement
1. Connect the power sensor to the computer as shown in Figure 5-2.
For a detailed description, refer to Chapter 3.5.1.1, "Simple USB Connection",
on page 14.
2. Start the R&S NRPV.
3. Execute zeroing:
Note: Turn off all measurement signals before zeroing. An active measurement
signal during zeroing causes an error.
a) Switch off the measurement signal.
R&S NRPV