Operating Concepts
R&S
®
NRPxxS(N)
35User Manual 1177.5079.02 ─ 10
Required equipment
R&S NRPxxS(N) power sensor
R&S NRPZKU cable or an R&S NRPZ5 sensor hub and an R&S NRPZK6 cable
to connect the sensor to the computer
Computer with installed:
R&S NRP Toolkit V 4.17 or higher
R&S Power Viewer version 9.2 or higher (refer to the operating manual of the
R&S Power Viewer for a description of the installation process)
If you want to use an android device like a tablet or a smartphone, use the R&S Power
Viewer Mobile. For details, see Chapter 5.6, "R&S Power Viewer Mobile",
on page 37.
Setup
Figure 5-3: Setup with the R&S Power Viewer
1 = Signal source
2 = R&S NRPxxS(N) power sensor
3 = Host interface connector
4 = R&S NRPZKU cable
5 = USB connector
6 = Computer with installed R&S Power Viewer
Incorrectly connecting/disconnecting the R&S NRPxxS(N) power sensors can damage
the power sensors or lead to erroneous results.
Ensure that you connect/disconnect your power sensor as described in Chapter 3,
"Preparing for Use", on page 11.
Starting a measurement
1. Connect the cables as shown in Figure 5-3.
For a detailed description, refer to Chapter 3.5.1.1, "Simple USB Connection",
on page 14.
2. Start the R&S Power Viewer.
3. Execute zeroing:
R&S Power Viewer