Operating concepts
R&S
®
NRT-Z14/44
30Getting Started 1443.8632.02 ─ 01
To ready the power sensor for measuring
1. Open the terminal program.
2. Send appl.
3. Wait for a response; see Chapter 5.3.2.3, "Behavior during booting",
on page 31.
4. Repeat step 2 and step 3 until the power sensor responds with:
@8E oper _____________________________________________
After receiving the response, you can make settings and query results.
To test the connection
1. Open the terminal program.
2. Enter the string test.
3. Confirm with the return key.
The power sensor responds:
@17 0 - This is a test string__________________
@18 1 - This is a test string__________________
@19 2 - This is a test string__________________
...
@37 97 - This is a test string__________________
@38 98 - This is a test string__________________
@39 99 - This is a test string__________________
See also Chapter 5.3.2.3, "Behavior during booting", on page 31.
To query the identification
► Send the id command.
The power sensor responds with its identification string.
Example: @6E Rohde&Schwarz NRT-Z14 V3.00 2021-12-01____
To query the last result
► Send rtrg.
The power sensor responds with:
@11 +2.1234E+01 +2.1530E+01 __avrl15500 ________
Remote control