Command reference
R&S
®
ZNA
1367User Manual 1178.6462.02 ─ 20
Suffix:
<Ch>
.
Channel number
Parameters:
<Pulse Duty Cycle> A value between 0 and 100.
Default unit: percent
Options: R&S ZNA<Freq>-B4<Pt> or R&S ZNA-B91
Manual operation: See "Fixed Duty Cycle/Duty Cycle" on page 641
[SENSe<Ch>:]PULSe:FXDCycle[:STATe]
<Boolean>
Defines whether the analyzer firmware shall enforce a fixed duty cycle for all pulse
generators.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1)
Fixed duty cycle for all pulse generators in channel <Ch>.
The pulses are defined using [SENSe<Ch>:]PULSe:PERiod
and [SENSe<Ch>:]PULSe:DUTYcycle.
OFF (0)
The generator-specific pulses are defined using [SENSe<Ch>:
]PULSe:PORT<Pt>:WIDTh and [SENSe<Ch>:]PULSe:
PORT<Pt>:DELay.
*RST: OFF
Options: R&S ZNA<Freq>-B4<Pt> or R&S ZNA-B91
Manual operation: See "Fixed Duty Cycle/Duty Cycle" on page 641
[SENSe<Ch>:]PULSe:GENerator:PERiod <Period for the generated pulses>
Suffix:
<Ch>
.
Parameters:
<Period for the
generated pulses>
[SENSe<Ch>:]PULSe:GENerator:SOURce <source for pulse generator>
Suffix:
<Ch>
.
Parameters:
<source for pulse
generator>
EXTA | EXTB | EXTC | EXTD | INT
SCPI command reference