Command reference
R&S
®
ZNA
1366User Manual 1178.6462.02 ─ 20
Suffix:
<Ch>
.
Parameters for setting and query:
<Gain> Gain factor for the internal low noise amplifier in dB
Range: 20 to 30
Increment: 5
Options: R&S ZNA-B302
Manual operation: See "Preamp. Gain" on page 478
7.3.14.23 [SENSe:]PULSe...
Pulsed measurements.
[SENSe<Ch>:]PULSe:ATIMe <Acquisition time>
Defines the duration of the data acquisition for pulse measurements in channel <Ch>.
Suffix:
<Ch>
.
Channel number
Parameters:
<Acquisition time> Data acquisition time
Default unit: s
Manual operation: See "Acq. Time/Fixed" on page 644
[SENSe<Ch>:]PULSe:COUPled[:STATe] <Boolean>
Enforces identical pulse generator settings for all ports.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> *RST: OFF (0)
Options: R&S ZNA<Freq>-B4<Pt> or R&S ZNA-B91
Manual operation: See "Coupled Settings" on page 642
[SENSe<Ch>:]PULSe:DUTYcycle <Percentage>
Defines the duty cycle of the generated pulses.
This setting only takes effect, if:
Internal Pulse Control is enabled ([SENSe<Ch>:]PULSe:INTernal[:STATe]
ON)
Fixed Duty Cycle is enabled ([SENSe<Ch>:]PULSe:FXDCycle[:STATe] ON)
Note that currently the exact value of the duty cycle can not be guaranteed. However,
you can be sure that the specified value is not exceeded.
SCPI command reference