644User Manual 1178.6462.02 ─ 20
PuMo In
If external Pulse Modulator Control is enabled, this setting allows you to select
between positive or negative polarity of the "PuMo In" signal.
Remote command:
[SENSe<Ch>:]PULSe:PORT<Pt>:EXTernal:INPut:INVerted[:STATe]
Trig. Out
Allows you to enable/disable pulse modulator control via the respective "Trig. Out".
Note that a Trigger Out that has been previously configured to indicate sweeps, seg-
ments or measurement points for this channel, cannot be used for pulse generation.
The corresponding table row is disabled in this case.
To change the trigger configuration, use the "Trigger Out Manager..." button to open
the Trigger Out Manager dialog.
Remote command:
TRIGger:CHANnel<Ch>:AUXiliary<n>:INTerval PGENerator | NONE
TRIGger:CHANnel<Ch>:AUXiliary<n>:STYPe POSitive | NEGative
Measurement settings
Sets up the "Pulse Profile" measurement for this channel (see Chapter 5.10.2, "Sweep
Type tab", on page 491).
Acq. Delay/Fixed
Defines the acquisition delay relative to the master pulse. Can only be set in synchron-
ized measurement mode.
Remote command:
[SENSe<Ch>:]PULSe:MEASdelay
Acq. Time/Fixed
Defines the duration of the data acquisition.
Remote command:
[SENSe<Ch>:]PULSe:ATIMe
Bandwidth/Fixed
Defines the resolution bandwidth.
Remote command:
[SENSe<Ch>:]BWIDth[:RESolution]
ALC Optimize
Enables/configures ALC optimization.
"Auto" (default)
The firmware decides whether speed optimization shall be performed
"On"
Optimization active
"Off"
Optimization inactive
Remote command:
SOURce<Ch>:POWer<Pt>:ALC:OPTimize
Mode softtool

Table of Contents