Command reference
R&S
®
ZNA
1372User Manual 1178.6462.02 ─ 20
This setting only takes effect, if:
Internal Pulse Control is enabled
Fixed Duty Cycle is disabled ([SENSe<Ch>:]PULSe:FXDCycle[:STATe] OFF)
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<pulse delay on pulse
modulator>
Pulse delay [s]
Default unit: s
Options: R&S ZNA<Freq>-B4<Pt> or R&S ZNA-B91
Manual operation: See "Pulse Delay/Pulse Width" on page 642
[SENSe<Ch>:]PULSe:PORT<Pt>:EXTernal:INPut:INVerted[:STATe]
<Boolean>
If External Pulse Control is used, this defines the polarity of the incoming generator
pulses at PuMo In/Out <Pt>.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<Boolean> *RST: OFF (0)
Options: R&S ZNA-B91
Manual operation: See "PuMo In" on page 644
[SENSe<Ch>:]PULSe:PORT<Pt>:EXTernal:OUTPut:INVerted[:STATe] <Boolean>
If pulse generator <Pt> is output to PuMo In/Out <Pt> (see [SENSe<Ch>:]PULSe:
PORT<Pt>:EXTernal:OUTPut[:STATe]), this defines the polarity of the outgoing
pulses.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<Boolean> *RST: OFF (0)
Options: R&S ZNA-B91
Manual operation: See "PuMo Out" on page 643