EasyManua.ls Logo

R&S ZNA26 - Page 1371

R&S ZNA26
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1371User Manual 1178.6462.02 ─ 20
Manual operation: See "Pulse Generator" on page 637
[SENSe<Ch>:]PULSe:INTernal[:STATe] <Boolean>
Switches between internal and external pulse control.
Suffix:
<Ch>
.
Channel number
Parameters:
<Boolean> ON (1)
Internal pulse control
OFF (0)
External pulse control
*RST: ON
Options: R&S ZNA-B91
Manual operation: See "Pulse Modulator Control" on page 643
[SENSe<Ch>:]PULSe:MEASdelay <Value>
Defines the measurement delay (acquisition delay) relative to the pulse start.
Suffix:
<Ch>
.
Channel number
Parameters:
<Value> Measurement delay [s]
Manual operation: See "Acq. Delay/Fixed" on page 644
[SENSe<Ch>:]PULSe:PERiod <pulse period for pulse modulators and trigger outs>
Defines the (common) pulse period of the internal pulse generators and modulators.
Suffix:
<Ch>
.
Channel number
Parameters:
<pulse period for
pulse modulators and
trigger outs>
Pulse period [s]
Default unit: s
Options: R&S ZNA-K7
Manual operation: See "Pulse Period" on page 640
[SENSe<Ch>:]PULSe:PORT<Pt>:DELay <pulse delay on pulse modulator>
Defines the port-specific delay of the pulses generated by the respective internal pulse
generator.