Command reference
R&S
®
ZNA
1373User Manual 1178.6462.02 ─ 20
[SENSe<Ch>:]PULSe:PORT<Pt>:EXTernal:OUTPut[:STATe] <Boolean>
If Internal Pulse Control is used, this enables/disables output of pulse generator <Pt>
at PuMo In/Out <Pt>.
Suffix:
<Ch>
.
Channel number
<Pt> Port number
Parameters:
<Boolean> *RST: OFF (0)
Options: R&S ZNA-B91
Manual operation: See "PuMo Out" on page 643
[SENSe<Ch>:]PULSe:PORT<Pt>:WIDTh
<pulse width for pulse modulators and
trigger outs>
Defines the port-specific width of the pulses generated by the respective internal pulse
generator.
This setting only takes effect, if:
●
Internal Pulse Control is enabled ([SENSe<Ch>:]PULSe:INTernal[:STATe]
ON)
●
Fixed Duty Cycle is disabled ([SENSe<Ch>:]PULSe:FXDCycle[:STATe] OFF)
Suffix:
<Ch>
.
Channel number
<Pt> VNA port number
Parameters:
<pulse width for pulse
modulators and
trigger outs>
Default unit: s
Options: R&S ZNA<Freq>-B4<Pt> or R&S ZNA-B91
Manual operation: See "Pulse Delay/Pulse Width" on page 642
[SENSe<Ch>:]PULSe:PORT<Pt>[:STATe] <Boolean>
Activates or deactivates the internal pulse modulator at VNA port <Pt>.
Suffix:
<Ch>
.
Channel number
<Pt> VNA port number
Parameters:
<Boolean>
Options: R&S ZNA<Freq>B4<Pt>
SCPI command reference