GUI reference
R&S
®
ZNA
337User Manual 1178.6462.02 ─ 20
"Add CW Mode"
Activates a new channel for the spectrum measurement. The "CW Frequency" plus
half the "Tone Distance" defines the center of the diagram.
"Spectrum = Marker"
This button is only available if the analyzer is equipped with software option
R&S ZNA-K1; see "Spectrum = Marker" on page 458. If there is no marker in the
active intermodulation trace, the center frequency of the related channel is used.
"Max IM Order"
Defines the width of the spectrum measurement.
Remote command:
[SENSe<Ch>:]FREQuency[:CW]
[SENSe<Ch>:]FREQuency:IMODulation:MSPectrum
[SENSe<Ch>:]FREQuency:IMODulation:SPECtrum:MORDer
[SENSe<Ch>:]FREQuency:IMODulation:SPECtrum[:STATe]
5.2.6.2 Intermodulation setup dialog
The intermodulation setup dialog is an instance of the Multi-channel setup dialog. It
allows you to set up one or more channels for intermodulation measurements on a
non-frequency converting DUT.