15Getting Started 1178.6456.02 ─ 18
Do not place the product next to heat-generating equipment such as radiators
or other products.
3.4.2 Mounting the product in a rack
To prepare the rack
1. Observe the requirements and instructions in "Setting up the product" on page 8.
2.
NOTICE! Insufficient airflow can cause overheating and damage the product.
Design and implement an efficient ventilation concept for the rack.
To mount the R&S ZNA in a rack
1. Use an adapter kit to prepare the R&S ZNA for rack mounting.
a) Order the rack adapter kit R&S ZZA-KN6 (order number 1332.4498.02).
b) Mount the adapter kit. Follow the assembly instructions provided with the
adapter kit.
2. Lift the R&S ZNA to shelf height.
3. Grab the handles and push the R&S ZNA onto the shelf until the rack brackets fit
closely to the rack.
4. Tighten all screws in the rack brackets with a tightening torque of 1.2 Nm to secure
the R&S ZNA in the rack.
To unmount the R&S ZNA from a rack
1. Loosen the screws at the rack brackets.
2. Remove the R&S ZNA from the rack.
3. If placing the R&S ZNA on a bench top again, unmount the adapter kit from the
R&S ZNA. Follow the instructions provided with the adapter kit.
3.5 Considerations for test setup
Cable selection and electromagnetic interference (EMI)
Electromagnetic interference (EMI) can affect the measurement results.
To suppress electromagnetic radiation during operation:
Use high-quality shielded cables, for example, double-shielded RF and LAN
cables.
Always terminate open cable ends.
Ensure that connected external devices comply with EMC regulations.
Considerations for test setup

Table of Contents