EasyManua.ls Logo

RayTek MARATHON MR - Traceability of Instrument Calibration

RayTek MARATHON MR
58 pages
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
58 Rev. E2 04/2011 Marathon MR
11.5 Traceability of Instrument Calibration
CU S T O M E R I N F R A R E D T H E R M OM E T E R S
Computerized cavity
(black body source)
500 to 2727°C
400 to 800°C
Medium Temp Secondary
(Working) Standard
400 to 800°C
Fluke Medium Temp
Primary Standard
760 to 2700°C
High Temp
Primary Standard
760 to 2700°C
High Temp Secondary
(Working) Standard
Working cavaties and plates
(black body sources)
25 to 500°C
-20 to 800°C
Low Temp Secondary
(Working) Standard
-20 to 800°C
Low Temp
Primary Standard
Daily check at 1000,
1700, and 2500°C
Cross check
Annual NIST Calibration
and historical data
results review
Annual NIST Calibration
and historical data
results review
Fluke Calibrated
blackbody cavaties
and plates
Fluke Working
Standards
Fluke Primary
Standards
National Metrology
Institutes
PTB
-50 to 800°C
ITS 90
NIST
800 to 2700°C
ITS 90
Fluke reserves the right to make changes to this traceability and methodes without notice.