ERROR INDICATORS
Air plug disconnected.
Arm cuff not applied
correctly.
Air is leaking from the
arm cuff.
Cuff over or under
inflated.
Batteries are depleted.
Insert the plug securely. Refer to
"APPLYING THE ARM CUFF".
Irregular heartbeats are
detected.
Remove the arm cuff. Wait 2-3 minutes and
then take another measurement.
Repeat the steps in "TAKING A
MEASUREMENT". If this error continues
to appear, contact your physician.
Movement during
measurement.
Remain still and do not talk during the
measurement.
Carefully read and repeat the steps in
"TAKING A MEASUREMENT".
Batteries are low.
Recommend to replace the batteries with
new ones ahead of time.
Refer to "BATTERY INSTALLATION".
Apply the arm cuff correctly. Refer to
"APPLYING THE ARM CUFF".
Replace the four batteries.
Refer to "BATTERY INSTALLATION".
Repeat measurement. Remain still and do
not talk during measurement. Refer to
"TAKING A MEASUREMENT".
If "E2" appears repeatedly, inflate the cuff
manually until it is 30 to 40 mmHg above
your previous measurement result. Refer
to "TAKING A MEASUREMENT".
Replace the cuff with a new one.
SYMBOL CAUSE CORRECTION