Model: HEM-780RELN3 (HEM-8700-WM)
Display: LCD Digital Display
Measurement Range: Pressure: 0 to 299 mmHg, Pulse: 40 to 180/min.
Accuracy/Calibration: Pressure: ±3mmHg or 2% of reading
Pulse: ±5% of reading
Inflation: Automatic by electric pump
Deflation: Automatic pressure release valve
Rapid Air Release: Automatic exhaust valve
Pressure Detection: Capacitive pressure sensor
Measurement Method: Oscillometric method
Pulse Wave Detection: Capacitance type pressure sensor
Power Source: 120V 60Hz AC adapter or 4 “AA” batteries
Battery Life: Approx. 1400 uses when used three times a day
with 4 new alkaline batteries
Operating Temperature 50ºF to 104ºF (10ºC to 40ºC) / 15 to 90 % RH
/Humidity:
Storage Temperature -4ºF to 140ºF (-20ºC to 60ºC) / 10 to 95% RH /
/Humidity/Air Pressure: 700 - 1060 hPa
Main Unit Weight: Approximately 14 1/8 oz. (400 g)
not including batteries
Main Unit Dimensions: Approx. 6 1/4" (l)
×
5" (w)
×
3 1/8" (h)
(157 mm
×
126 mm
×
78 mm)
Cuff Size: Approximately 6" (w) (152 mm)
Cuff tube 23 5/8" (l) (600 mm)
Cuff Circumference: Fits arm circumferences 9" to 17"
(220 mm to 420 mm)
Memory: Up to 84 per user
Contents: Main Unit, Easy Wrap Cuff, AC Adapter,
Instruction Manual, 4 “AA” batteries and Storage Case
NOTE: These specifications are subject to change without notice.
SPECIFICATIONS