39
ERROR CODES
DSD04
Solve problem:
EEEEE = offrange or conversion
error
Internal A/D converter error. Change the electronic board/ check the
load cell/check the maximum weight parameter
NO CAL = no calibration present
There isn' t any calibration. Start a new calibration
BAT LOW = low battery
Battery discharged. Change the battery
LIN ERR = error in the linearization
procedure
Too few divisions or wrong linearization points . Make a new
linearization.
OVLLOAD = weight over the
maximum capacity
Check the load cell