XCell
TM
ATF System with C410:V4B Controller User Guide
72
Input/ Output List
Tag Name
Description
PLC Tag Address
Engineering
Valve Fail
Position
Real Time
Trend
Range
Decimal Places
Units
PP1
System Stop
%I0.0
N/A
N/A
N/A
N/A
N/A
PP2
System Stop
%I0.1
N/A
N/A
N/A
N/A
N/A
-
Spare
N/A
N/A
N/A
N/A
N/A
N/A
-
Spare
N/A
N/A
N/A
N/A
N/A
N/A
-
Spare
N/A
N/A
N/A
N/A
N/A
N/A
-
Spare
N/A
N/A
N/A
N/A
N/A
N/A
-
Spare
N/A
N/A
N/A
N/A
N/A
N/A
-
Spare
N/A
N/A
N/A
N/A
N/A
N/A
-
Spare
N/A
N/A
N/A
N/A
N/A
N/A
-
Spare
N/A
N/A
N/A
N/A
N/A
N/A
-
Spare
N/A
N/A
N/A
N/A
N/A
N/A
-
Spare
N/A
N/A
N/A
N/A
N/A
N/A
-
Spare
N/A
N/A
N/A
N/A
N/A
N/A
-
Spare
N/A
N/A
N/A
N/A
N/A
N/A
PL1
System Stop Indicator
%Q0.0
N/A
N/A
N/A
N/A
N/A
HN1
System Fault Warning Horn
%Q0.1
N/A
N/A
N/A
N/A
N/A
SOL1
Pump Inflate/Deflate Solenoid
%Q0.2
N/A
N/A
N/A
N/A
N/A
CR1
Pump Controller Status Relay
#1
%Q0.3
N/A
N/A
N/A
N/A
N/A
CR2
Pump Controller Status Relay
#2
%Q0.4
N/A
N/A
N/A
N/A
N/A
CR3
Pump Interlock Relay #3
%Q0.5
N/A
N/A
N/A
N/A
N/A
CR4
Pump Interlock Relay #4
%Q0.6
N/A
N/A
N/A
N/A
N/A
PL2
System Stop Indicator
%Q0.7
N/A
N/A
N/A
N/A
N/A
-
Spare
N/A
N/A
N/A
N/A
N/A
N/A
-
Spare
N/A
N/A
N/A
N/A
N/A
N/A
-
Spare
N/A
N/A
N/A
N/A
N/A
N/A
RAW_P2_PV_1
Pump Pressure
%IW96
Note 1
1
PSI
N/A
Y
-44.7
RAW_PRV1_PV_1
Inflation Pressure
%IW98
Note 1
1
PSI
N/A
Y
0 - 40
RAW_PRV2_PV_1
Deflation Pressure
%IW100
Note 1
1
PSI
N/A
Y
-15
RAW_AIW10_PV_1
Spare
N/A
N/A
N/A
N/A
N/A
N/A
RAW_P3_PV_1
Optional Pressure 3
%IW104
Note 1
N/A
PSI
N/A
Y
RAW_P4_PV_1
Optional Pressure 4
%IW106
Note 1
N/A
PSI
N/A
Y
RAW_P5_PV_1
Optional Pressure 5
%IW108
Note 1
N/A
PSI
N/A
Y
RAW_W1_PV_1
Optional Weight
%IW110
Note 1
N/A
LBS
N/A
Y
RAW_PRV1_CV_1
Inflation Pressure
%QW128
0-100
1
%
FC
Y
RAW_PRV2_CV_1
Deflation Pressure
%QW130
0-100
1
%
FO
Y
RAW_PV1_CV_1
Inflation Pressure Proportional
%QW132
0 - 40
1
PSI
N/A
Y
RAW_PV2_CV_1
Deflation Pressure Proportional
%QW134
-15
1
PSI
N/A
Y
Note 1: Depends on calibration

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the REPLIGEN XCell ATF Series and is the answer not in the manual?

REPLIGEN XCell ATF Series Specifications

General IconGeneral
BrandREPLIGEN
ModelXCell ATF Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals