Spectrum® KR2i
/ KMPi TFF Systems
USER GUIDE AND OPERATING INSTRUCTIONS
Spectrum® KrosFlo® Research 2i TFF System
Spectrum® KrosFlo® MiniKros Pilot i TFF System

Table of Contents

Questions and Answers

  • R
    rachaellopezAug 9, 2025
    Why is my REPLIGEN Spectrum KMPi Laboratory Equipment process ending early or late in DV or CF auto modes?
    • K
      Kelly McguireAug 9, 2025
      If your REPLIGEN Laboratory Equipment process is ending prematurely or belatedly in DV or CF auto modes, the input for the holdup volumes may be incorrect. Ensure you enter the correct feed holdup and permeate holdup volumes, and accurately specify whether they are empty or full when initiating the process.
  • T
    Tracy WillisAug 18, 2025
    What to do if REPLIGEN Spectrum KMPi process ends early when using Conductivity or UV for auto modes?
    • A
      adam49Aug 18, 2025
      If the REPLIGEN Laboratory Equipment process terminates prematurely when utilizing Conductivity or UV for auto modes, it may be due to the sensor not being completely filled with the appropriate liquid. The conductivity sensor should be filled with the same buffer as the sample at the start of the process. Likewise, the UV sensor should be zeroed using the same buffer the sample is in.
  • T
    Tina McguireAug 29, 2025
    How to fix wrong pressure sensor readings on REPLIGEN Laboratory Equipment?
    • M
      Molly HarringtonAug 29, 2025
      If the pressure sensor readings on your REPLIGEN Laboratory Equipment are incorrect, displaying either -9.9 psi or >35psi when there is no pressure applied, it may indicate a malfunctioning pressure sensor octopus cable or a broken pressure sensor. You should replace the Octopus Cable and/or the pressure sensor.

Summary

Safety Precautions

Explanation of Symbols

Explains warning symbols and their meanings, including Danger, Warning, and Caution.

Specifications

KR2 i Specifications

Details technical specifications for the KR2i system, including output, input, construction, and environment.

KMPi Specifications

Details technical specifications for the KMPi system, covering output, input, construction, and environment.

System Configuration and Major Components

Setup and Operation

KF Comm Pump Control Interface

Details on using the KF Comm software for system control.

Hardware Setup

Configuring system hardware parameters and settings.

Automated Mode Control Strategy

Defining set-points and control mechanisms for automated processes.

Diafiltration Set Points

Options for controlling diafiltration endpoints.

Concentration Set Points

Options for controlling concentration endpoints.

Pressure Calibration Procedure

Steps for calibrating pressure sensors for accurate readings.

Alarms and Stops Configuration

Setting up safety alarms and operational stops for the system.

System Functionality

Detailed Pump Mode Operation

In-depth explanation of various pump operating modes.

Concentration Mode

Detailed operation of the Concentration Mode.

Concentration;Diafiltration Mode

Operation of the Concentration/Diafiltration Mode.

Constant Feed Concentration Mode

Operation of the Constant Feed Concentration (CFC) mode.

Ctrl Mode Operation

Operation of the Control Mode.

Pump Head Setup and Operation

KR2 i Pump Head Operation

Procedures for installing and removing KR2i pump heads.

KMPi Pump Head Operation

Procedures for installing and operating KMPi pump heads.

Auxiliary Component Setup and Operation

Auxiliary Pumps Setup

Connecting and configuring auxiliary pumps.

Automatic Backpressure Valve

Setup and operation of automatic backpressure valves.

KONDUIT System Integration

Integrating the KONDUIT unit with the TFF system.

KF Comm Software

Installing and Updating KF Comm Software and Firmware

Guide to installing and updating system software and firmware.

Troubleshooting

Common Symptoms and Remedies

Identifies common issues and provides solutions.

REPLIGEN Spectrum KMPi Specifications

General IconGeneral
ModelSpectrum KMPi
ManufacturerRepligen
ApplicationsTangential flow filtration
Operating PressureUp to 100 psi
Pressure RatingUp to 6 bar (87 psi)
Membrane AreaVariable, depending on module selection
MaterialStainless steel
Data LoggingIntegrated data logging capabilities
ConnectivityEthernet
DimensionsVaries depending on configuration
WeightVaries depending on configuration
CompatibilityCompatible with various TFF membranes