29 400-12355-000 Rev. 03
KR2i / KMPi TFF Systems
Setup and Operation
Diafiltration 2 Set Point (Diaf 2 used in CDDC mode for second wash buffer):
1. DV (Diafiltration Volumes)
a. End-point control when permeate scale reaches target DV based on weight of feed
scale and holdup loop
b. Example Application: diafiltration of virus or protein
2. Cond 1
a. End-point control when conductivity reaches setpoint of Conductivity Sensor 1
b. Example Application: diafiltration of protein or oligo
3. UV
a. Semi-quantitative end-point control when UV reaches target AU
b. Example Application: diafiltration of latex particle with free IgG measurement on
permeate; offline confirmation needed for verification
Concentration 1 Set Point:
1. CF (Concentration Factor)
a. End-point control when retentate and permeate scale reaches target CF based on
weight of feed scale and holdup loop
b. Example Application: concentration of virus or protein
2. UV
a. Semi-quantitative end-point control when UV reaches target AU, Application
dependent, offline confirmation needed for verification
b. Example Application: concentration of oligo
3. UV2
1. Semi-quantitative end-point control when UV reaches target AU, Application
dependent, offline confirmation needed for verification
2. Example Application: concentration of oligo
Concentration 2 Set Point (Conc 2 used for final concentration in CDC):
1. CF (Concentration Factor)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the REPLIGEN Spectrum KMPi and is the answer not in the manual?

REPLIGEN Spectrum KMPi Specifications

General IconGeneral
BrandREPLIGEN
ModelSpectrum KMPi
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals