XCell
TM
ATF System with C410:V4B Controller User Guide
79
d. Set InputPage3.P2_P2_Capture = 1 until the OutputPage4.P2_Cal_Enable bit is
received.
i. Wait for the “RequestedDataPage” to be equal to 4; since cycling through
output data pages is recommended.
ii. Examine OutputPage4.P2_ P2_Capture. When true…
e. Reset InputPage3.P2_P2_Capture = 0; the maximum capture bit is momentarily
asserted.
6) Assert accept bit:
a. While “SendingDataPage” is still equal to 3…
b. Set InputPage3.P2_Accept = 1 until the OutputPage4.P2_Accept bit is received.
i. Wait for the “RequestedDataPage” to be equal to 4; since cycling through
output data pages is recommended.
ii. Examine OutputPage4.P2_Accept. When true…
c. Reset InputPage3.P2_Accept = 0.
d. Controller performs the linear scaling calculation and resets the controller’s
internal calibration enable bit.
7) Clean up: After all calibration activities are complete, be sure to perform the following:
a. Reset parameters that were set during the calibration process.
b. Set “SendingDataPage” = 0 to prepare for next ATF run commands.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the REPLIGEN XCell ATF10 and is the answer not in the manual?

REPLIGEN XCell ATF10 Specifications

General IconGeneral
BrandREPLIGEN
ModelXCell ATF10
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals