EasyManuals Logo

ResMed AirFit N30 Disinfection Guide

ResMed AirFit N30
3 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
English 1
Disinfection guide Clinical use only
This guide is intended for multi-patient use of the AirFit
N30 nasal mask in a sleep lab, clinic or hospital. If you use the mask as a single user in the home, refer to the User Guide for cleaning instructions. The
processor is responsible for ensuring that reprocessing is completed in accordance with ResMed’s validated procedures.
Mask component
1
Cleaning
Validated number of cycles
Mild Alkaline, Anionic Detergent e.g. Alconox Neodisher Mediclean Forte
Thermal Disinfection
Manual Washer-Disinfector
Cushion 30
Standard frame
30
QuietAir™ frame
2
Split-band headgear 30
1
May not be available in all regions. For full details regarding the correct use of this mask, please refer to the User Guide. For a list of available replacement parts, check the Product Guide on ResMed.com
2
Not intended for multi-patient re-use.
A Split-band headgear
1 Buckle
2 Buttonhole
3 End cap
B Cushion
C QuietAir™ / Standard frame
4 Stabiliser
5 QuietAir vent
6 Vent
7 Mask tubing

Other manuals for ResMed AirFit N30

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed AirFit N30 and is the answer not in the manual?

ResMed AirFit N30 Specifications

General IconGeneral
Cushion MaterialSilicone
TypeNasal Mask
Sizes AvailableSmall, Medium
CompatibilityCompatible with most CPAP machines
Headgear TypeAdjustable
Vent TypeQuietAir vent technology
Recommended CleaningDaily with mild soap and water; air dry