English 1
Disinfection guide Clinical use only
This guide is intended for multi-patient use of the AirFit
™
N30 nasal mask in a sleep lab, clinic or hospital. If you use the mask as a single user in the home, refer to the User Guide for cleaning instructions. The
processor is responsible for ensuring that reprocessing is completed in accordance with ResMed’s validated procedures.
Mask component
1
Cleaning
Validated number of cycles
Mild Alkaline, Anionic Detergent e.g. Alconox Neodisher Mediclean Forte
Thermal Disinfection
Manual Washer-Disinfector
• Cushion 30
• Standard frame
30
• QuietAir™ frame
2
– – –
• Split-band headgear 30
1
May not be available in all regions. For full details regarding the correct use of this mask, please refer to the User Guide. For a list of available replacement parts, check the Product Guide on ResMed.com
2
Not intended for multi-patient re-use.
A Split-band headgear
1 Buckle
2 Buttonhole
3 End cap
B Cushion
C QuietAir™ / Standard frame
4 Stabiliser
5 QuietAir vent
6 Vent