Cleaning and maintenance
136
Servicing
WARNING
Inspection and repair should only be performed by an authorised agent. Under no circumstances
should you attempt to service or repair the device yourself. Failure to do so could void your Astral
device warranty, damage the Astral device or result in possible injury or death.
Note: Retain the original packaging of the Astral device for use when shipping to/from an authorised ResMed Service
Centre.
Maintenance Timetable
The Astral device should be serviced by an authorised ResMed Service Centre according to the following
schedule. The Astral device is intended to provide safe and reliable operation provided that it is operated
and maintained in accordance with the instructions provided by ResMed. As with all electrical devices, if
any irregularity becomes apparent, you should exercise caution and have the device inspected by an
authorised ResMed Service Centre.
With regular servicing, the expected service life of an Astral device is 8 years.
Servicing schedule from the date of first use:
Recommended service interval Conducted by Instructions
Every six months
Personnel who have been trained in the
use of Astral
Replacement of the air filter (replace
earlier if dirty).
Replacement of Single or Double limb
circuit adapters if used.
Two years Qualified technician
Two year Preventative maintenance.
Replacement of the internal battery and
FiO
2
sensor if fitted.
35,000 hours Qualified technician
Pneumatic block Preventative
maintenance.
Helpful hint!
If Service Reminders are enabled you will get a notification on the device when two year preventative
maintenance is nearly due.
Internal Battery
The expected life of the internal battery is two years. The internal battery should be replaced every two
years or sooner when there is a noticeable reduction in usage time when fully charged. During storage
ensure that internal battery is recharged once every six months.
Helpful hint!
If Service reminders are enabled you will get a notification on the device when there is a measurable
reduction in battery usage time.

Other manuals for ResMed Astral Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed Astral Series and is the answer not in the manual?

ResMed Astral Series Specifications

General IconGeneral
TypeVentilator
Displaycolor LCD
ConnectivityUSB
EPAP0 to 25 cm H2O
Rise TimeAdjustable
PEEP0 to 25 cm H2O
FiO2 MonitoringYes, with optional oxygen module
Battery BackupInternal and external battery options
Oxygen CapabilityYes, with optional oxygen module
AlarmsLow battery
Pressure Support0 to 30 cm H2O