English 1
H5i
™
disinfection guide
This disinfection guide is intended for multipatient use of the H5i in a sleep lab, clinic, hospital or at a health
care provider. If you are using the H5i as a single user in the home, refer to the Welcome Guide for cleaning
instructions.
This guide describes ResMed’s recommended and validated procedures for cleaning and disinfection of the
humidifier. However, the steps for disinfection vary regionally. As a result, each healthcare facility should consult its
own procedures before carrying out those within this guide.
GENERAL WARNINGS AND CAUTIONS
• ResMed cannot give any assurance that deviations from the procedures listed in this guide, and
their effect on the performance of the product, will be acceptable.
• When using detergents, disinfectants or sterilization agents, always follow the manufacturer’s
instructions.
• Beware of electrocution. Do not immerse the device, power supply or power cord in water. Always
unplug the device before cleaning and be sure that it is dry before plugging back in.
Parts suitable for disinfection or sterilization
Part
High level thermal
disinfection
High level chemical
disinfection
Sterilization
H5i cleanable water tub
(including tub lid, plate and base)
Air outlet
English