9Working with other optional accessories
English
To add supplemental oxygen
1 Unlock the low flow oxygen inlet on the back of the device by pushing up on the locking
clip.
2 Insert one end of the oxygen supply tubing into the oxygen connector port. The tubing
automatically locks into place.
3 Attach the other end of the oxygen supply tubing to the oxygen supply.
4 Start ventilation.
5 Turn on the oxygen source and adjust to the desired flow rate.
To remove supplemental oxygen
Before you remove supplemental oxygen from the device, ensure the oxygen supply has
been turned off.
1 Unlock the low flow oxygen inlet on the back of the device by pushing up on the locking
clip.
2 Remove the oxygen supply tubing from the oxygen connector port.
1
2
1
2

Other manuals for ResMed stellar series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the ResMed stellar series and is the answer not in the manual?

ResMed stellar series Specifications

General IconGeneral
Data StorageSD card
Device TypeVentilator
Modes of VentilationCPAP, S, S/T, T
Power SupplyAC 100-240V, 50/60Hz
Operating Temperature5°C to 40°C
Operating Humidity10% to 95% (non-condensing)
Operating AltitudeUp to 2500 meters
Sound Level30 dBA
HumidificationIntegrated heated humidifier
ConnectivityWi-Fi